FLUKE-287 CALIBRATED Fluke, FLUKE-287 CALIBRATED Datasheet - Page 55

MULTIMETER, DIGITAL, HANDHELD

FLUKE-287 CALIBRATED

Manufacturer Part Number
FLUKE-287 CALIBRATED
Description
MULTIMETER, DIGITAL, HANDHELD
Manufacturer
Fluke
Datasheet

Specifications of FLUKE-287 CALIBRATED

Dmm Type
Handheld
Voltage Measuring Range Dc
50 MV To 50 V
Voltage Measuring Range Ac
500 V To 1000 V
Current Measuring Range Dc
500µA To 10A
Current Measuring Range Ac
500µA To 10A
Lead Free Status / RoHS Status
na
To measure duty cycle, position the rotary switch on one of the
functions allowing frequency measurements shown in Figure 23.
Press the softkey labeled Menu and move the menu selector to
the menu item labeled Hz,%,ms. Next press the softkey labeled
%.
As shown in Figure 26, the duty cycle percentage is shown in the
primary display while the signal frequency appears in the
secondary display. The mini-measurement display indicates the
0
Menu
8:10pm
Figure 26. Duty Cycle Display
100
49.75
Save
59.756
200
123.45 VAC
300
400
%
Hz
Duty Cycle
06/13/07
Auto Range
Setup
500 VAC
est24.eps
volts or amps value of the input signal. The bar graph tracks the
volts or amps value of the signal and not the duty cycle value.
The pulse polarity is displayed to the right of the duty cycle value.
J indicates a positive pulse and K indicates a negative pulse.
To change the polarity being measured, press the softkey
labeled J K. The polarity indicator changes to the opposite
polarity.
For 5 V logic signals, use the 5 V dc range. For 12 V switching
signals in automobiles, use the 50 V dc range. For sine waves,
use the lowest ac or dc range that does not result in multiple
triggering. A manually-selected lower input range will often
measure better than the AUTO-selected input range.
Measuring Pulse Width
The pulse width function measures the amount of time a signal is
high or low, as shown in Figure 27. The measured waveform
must be
The meter measures pulse width from 0.025 ms to 1250.0 ms
ranges.
To measure pulse width, position the rotary switch to one of the
functions allowing frequency measurements shown in Figure 23.
Press the softkey labeled Menu and move the menu selector to
the menu item labeled Hz,%,ms. Next, press the softkey labeled
ms.
periodic;
its pattern must repeat at equal time intervals.
Making Measurements
45