E5062A/150/015 Agilent Technologies Test Equipment, E5062A/150/015 Datasheet

no-image

E5062A/150/015

Manufacturer Part Number
E5062A/150/015
Description
E5061A ENA-L NETWORK ANALYZER, 2-PORT, 300 KHZ TO 3.0 GHZ, TR TEST SET
Manufacturer
Agilent Technologies Test Equipment
Type
Networkr
Datasheet

Specifications of E5062A/150/015

Frequency Range
300 kHz-3.0 GHz
Number Of Ports
2
Special Features
T/R
Agilent
ENA-L RF Network Analyzers
E5061A, 300 kHz to 1.5 GHz
E5062A, 300 kHz to 3 GHz
Data Sheet

E5062A/150/015 Summary of contents

Page 1

Agilent ENA-L RF Network Analyzers E5061A, 300 kHz to 1.5 GHz E5062A, 300 kHz to 3 GHz Data Sheet ...

Page 2

Definitions All specifications apply over a 23 °C ±5 °C range (unless otherwise stated) and 90 minutes after the instrument has been turned on. Specification (spec.): Warranted performance. Specifications include guardbands to account for the expected statistical performance distribution, measurement ...

Page 3

Corrected system performance The specifications in this section apply for measurements made with the Agilent E5061A/E5062A network analyzer with the following conditions: • No averaging applied to data • Environmental temperature of 23 °C ±5 °C, with less than 1 ...

Page 4

Table 1-2 Corrected system performance with Type-N 50 Ω connectors, 85032F calibration kit, full 2-port calibration Network analyzer: E5061A/E5062A, calibration kit: 85032F (Type-N, 50 Ω), calibration: full 2-port IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature ...

Page 5

Table 1-3 Corrected system performance with Type-N 50 Ω connectors, 85032F calibration kit, enhanced response calibration Network analyzer: E5061A/E5062A, calibration kit: 85032F (Type-N, 50 Ω) calibration: enhanced response IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature ...

Page 6

Table 1-4 Corrected system performance with Type-N 75 Ω connectors 85036E calibration kit, full 2-port calibration Network analyzer: E5061A/E5062A, calibration kit: 85036E (Type-N, 75 Ω), calibration: full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature ...

Page 7

Table 1-5 Corrected system performance with Type-N 75 Ω connectors 85036E calibration kit, enhanced response calibration Network analyzer: E5061A/E5062A, calibration kit 85036E (Type-N, 75 Ω), calibration: enhanced response IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature ...

Page 8

Uncorrected system performance Table 1-6 Uncorrected system performance (correction: off, 23 °C ±5 °C) Description Directivity Source match Load match Transmission tracking Reflection tracking Test port output (source) Table 1-7 Test port output frequency Description Range E5061A E5062A Resolution Source ...

Page 9

Test port output (source) Table 1-8 Test port output power Description 1 Level accuracy (at 23 °C ±5 °C) 300 kHz to 3 GHz Level linearity (at 23°C ±5°C) 300 kHz to 3 GHz Range (standard) 300 kHz to 3 ...

Page 10

Test port input Table 1-10 Test port input levels Description Maximum test port input level 300 kHz to 3 GHz Damage level 300 kHz to 3 GHz 1 Crosstalk 300 kHz to 3 GHz Table 1-11 Test port input (trace ...

Page 11

Table 1-13 Test port input (dynamic accuracy) Accuracy of the test port input power reading is relative to –10 dBm reference input power level. Specification Magnitude Typical Phase 11 ...

Page 12

Table 1-14 Test port input (group delay Description Aperture (selectable) Maximum aperture Minimum delay Accuracy The following graph shows group delay accuracy with Type-N full 2-port calibration and bandwidth. Insertion loss is assumed to be < ...

Page 13

General information Table 1-15 System bandwidths Description IF bandwidth settings Range Table 1-16 Front panel information Description RF connectors Type Display Size Resolution 1. Valid pixels are 99.99% and more. Below 0.01% of fixed points of black, blue, green or ...

Page 14

Table 1-17 Rear panel information Description External trigger connector Type Input level Pulse width Polarity External reference signal input connector Type Input frequency Input level Internal reference signal output connector Type Output frequency Signal type Output level Output impedance VGA ...

Page 15

Table 1-18 EMC and safety Description EMC ISM 1-A ICES/NMB-001 N10149 Safety ISM 1-A LR95111C ® General characteristics European Council Directive 89/336/EEC EN / IEC 61326-1:1997+A1:1998 CISPR 11:1997+A1:1999 / EN 55011:1998+A1:1999 Group 1, Class A IEC 61000-4-2:1995 / EN 61000-4-2:1995+A1:1998 ...

Page 16

Table 1-19 Analyzer environment and dimensions Description Operating environment Temperature Error-corrected temperature range Humidity Altitude Vibration Non-operating storage environment Temperature Humidity Altitude Vibration Dimensions Weight Figure 1-2. Dimensions (front view, in millimeters) 16 General characteristics +5 °C to +40 °C ...

Page 17

Figure 1-3. Dimensions (rear view, in millimeters) Figure 1-4. Dimensions (side view, in millimeters) 17 ...

Page 18

Measurement throughput summary Table 1-20 Typical cycle time for measurement completion Start 1 GHz, stop 1.2 GHz, 30 kHz IF bandwidth Uncorrected 2-port cal Start 300 kHz, stop 1.5 GHz, 30 kHz IF bandwidth Uncorrected 2-port cal Start 300 kHz, ...

Page 19

Table 1-22 Data transfer time Number of points 2 SCPI over GPIB REAL 64 ASCII 2 SCPI over 100 Mbps LAN (telnet) REAL 64 ASCII 2 SCPI over 100 Mbps LAN (SICL-LAN) REAL 64 ASCII 2 COM (program executed in ...

Page 20

Measurement capabilities Number of measurement channels Number of display windows Number of traces Measurement choices Measurement parameter conversion Data formats Data markers Marker functions Marker search Marker-to functions Search range Tracking Fault location functions (Option E5061A/E5062A-100) Transformation to distance and ...

Page 21

Trace functions Display data Trace math Title Autoscale Electrical delay Phase offset Statistics Data accuracy enhancement Measurement calibration Calibration types available Response Response and isolation Enhanced response One-port calibration Full 2-port calibration (Option E5061A/E5062A-250/275) Interpolated error correction Velocity factor Reference ...

Page 22

Storage Internal hard disk drive File sharing Disk drive Screen hardcopy System capabilities Familiar graphical user interface Limit lines 22 Store and recall instrument states, calibration data, and trace data on 10 GB, minimum, internal hard drive. Trace data can ...

Page 23

Automation GPIB SCPI X COM Methods Internal analyzer execution Controlling via GPIB LAN Standard conformity Protocol Function Web Resources For additional literature and product information about the Agilent ENA-L visit: www.agilent.com/find/ena Internal X X Applications can be developed in a ...

Page 24

Agilent Email Updates www.agilent.com/fi nd/emailupdates Get the latest information on the products and applications you select. Agilent Direct www.agilent.com/fi nd/agilentdirect Quickly choose and use your test equipment solutions with confi dence. Agilent Open www.agilent.com/fi nd/open Agilent Open simplifi es the ...