Specifications of E5062A/150/015

Frequency Range
300 kHz-3.0 GHz
Number Of Ports
2
Special Features
T/R
Corrected system performance
The specifications in this section apply for measurements made with the
Agilent E5061A/E5062A network analyzer with the following conditions:
• No averaging applied to data
• Environmental temperature of 23 °C ±5 °C, with less than 1 °C deviation
• Response and isolation calibration not omitted
Table 1-1
Description
System dynamic range
300 kHz to 1 MHz, IF bandwidth = 3 kHz
1 MHz to 3 GHz, IF bandwidth = 3 kHz
300 kHz to 1 MHz, IF bandwidth = 10 Hz
1 MHz to 3 GHz, IF bandwidth = 10 Hz
System dynamic range; specification and measurement example
1. The test port dynamic range is calculated as the difference between the test port rms noise floor
2. Applicable to the units with serial prefix MY442 and above .
from the calibration temperature
and the source maximum output power. The effective dynamic range must take measurement
uncertainty and interfering signals into account.
160
150
140
130
120
110
100
90
80
3.0E+05
5.0E+08
System dynamic range
1.0E+09
System dynamic range
Test frequency [/Hz]
1.5E+09
Specification
90 dB
95 dB
115 dB
120 dB
1 2
2.0E+09
E5061A/62A spec.
2.5E+09
SPD
130 dB
3