E5062A/150/015 Agilent Technologies Test Equipment, E5062A/150/015 Datasheet - Page 12

no-image

E5062A/150/015

Manufacturer Part Number
E5062A/150/015
Description
E5061A ENA-L NETWORK ANALYZER, 2-PORT, 300 KHZ TO 3.0 GHZ, TR TEST SET
Manufacturer
Agilent Technologies Test Equipment
Type
Networkr
Datasheet

Specifications of E5062A/150/015

Frequency Range
300 kHz-3.0 GHz
Number Of Ports
2
Special Features
T/R
12
Table 1-14
Description
Aperture (selectable)
Maximum aperture
Minimum delay
Accuracy
The following graph shows group delay accuracy with Type-N full 2-port calibration and a 10 Hz IF bandwidth. Insertion loss is assumed to be < 2 dB.
Group delay (typical)
In general, the following formula can be used to determine the accuracy, in seconds, of specific group delay measurement:
±phase accuracy (deg)/[360 x aperture (Hz)]
1. Group delay is computed by measuring the phase change within a specified step (determined by the
frequency span and the number of points per sweep).
Test port input (group delay
Specification
(frequency span)/(number of points –1)
25% of frequency span
1
)
Supplemental information
Limited to measuring no more than
180° of phase change within the minimum aperture.
See graph below