E5062A/150/015 Agilent Technologies Test Equipment, E5062A/150/015 Datasheet - Page 20

no-image

E5062A/150/015

Manufacturer Part Number
E5062A/150/015
Description
E5061A ENA-L NETWORK ANALYZER, 2-PORT, 300 KHZ TO 3.0 GHZ, TR TEST SET
Manufacturer
Agilent Technologies Test Equipment
Type
Networkr
Datasheet

Specifications of E5062A/150/015

Frequency Range
300 kHz-3.0 GHz
Number Of Ports
2
Special Features
T/R
20
Measurement capabilities
Number of measurement channels
Number of display windows
Number of traces
Measurement choices
Measurement parameter conversion
Data formats
Data markers
Marker functions
Marker search
Marker-to functions
Search range
Tracking
Fault location functions (Option E5061A/E5062A-100)
Transformation to distance and time domain
LXI compliance
Source control
Measured number of points per sweep
Sweep type
Segment sweep
Sweep trigger
Power
User definable from 2 to 1601.
Linear sweep, segment sweep, log sweep and power sweep.
Define independent sweep segments. Set number of points, test port power levels,
IF bandwidth, delay time, sweep time and sweep mode independently for each segment.
Set to continuous, hold, or single, sweep with internal, external, manual,
or bus trigger.
Set source power from -5 dBm (-45 dBm for option E5061A/E5062A-1E1/250/275)
to 10 dBm. The power slope function compensates source power level error.
Up to 4 independent measurement channels. A measurement channel is coupled
to stimulus response settings including frequency, IF bandwidth, power level,
and number of points.
Each measurement channel has a display window. Up to 4 display windows (channels)
can be displayed.
4 data traces and 4 memory traces per channel
Option E5061A/E5062A-150/175: S
Option E5061A/E5062A-250/275: S
Available to convert S-parameters into reflection impedance, transmission impedance,
reflection admittance, transmission admittance, and 1/S.
Log magnitude, linear magnitude, phase, expanded phase, positive phase,
group delay, SWR, real, imaginary, Smith chart, polar.
10 independent markers per trace. Reference marker available
for delta marker operation. Smith chart format includes 5 marker formats:
linear magnitude/phase, log magnitude/phase, real/imaginary,
R + jX, and G + jB. Polar chart format includes 3 marker formats:
linear magnitude/phase, log magnitude/phase, and real/imaginary.
Max value, min value, multi-peak, multi-target, peak, peak left, peak right, target,
target left, target right, and width parameters with user-defined bandwidth values.
Set start, stop, center to active marker stimulus value; set reference to active
marker response value; set electrical delay to group delay at active marker.
User definable.
Performs marker search continuously or on demand.
Selectable transformation type from bandpass, lowpass impulse, lowpass step.
Selectable window from maximum, normal and minimum.
Class C (only applies to units that are shipped with firmware revision A.03.00 or later).
11
11
, S
, S
21
21
, S
12
, S
22