8845A/SU 120V Fluke Electronics, 8845A/SU 120V Datasheet - Page 84

PREC MLTMTR DMM BNCH 6.5 DIG RES

8845A/SU 120V

Manufacturer Part Number
8845A/SU 120V
Description
PREC MLTMTR DMM BNCH 6.5 DIG RES
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8845A/SU 120V

Includes
Test Leads
Style
Bench
Display Digits
6.5
Display Type
VFD, Dual
Display Count
Varies
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Memory, RS-232
Ranging
Manual
Response
Average
Lead Free Status / RoHS Status
Contains lead / RoHS compliant by exemption
Other names
2675292
614-1001
8845A/8846A
Users Manual
4-16
Checking Diodes
None
The diode function sends a current through a semiconductor junction while the Meter
measures the voltage drop across the junction (or junctions). Measurements are displayed
in the 10 V range at a relatively fast measurement rate. "OPEN" is displayed for voltages
10 % above the compliance voltage setting. Typical junction voltage drop on good
junctions are 0.3 to 0.8 volts. If enabled, the beeper will make a short beep when a good
junction is detected. Shorted diodes will indicate a significantly lower voltage.
With the higher compliance voltage (up to 10 V), the Meter’s diode test function can test
zeners up to 10 volts, diode stacks and LEDs. The selectable current and maximum
voltage allows you to tailor the diode test to the expected voltage for the junction being
tested.
To check a diode:
1. Press S twice on the 8846A or D once on the 8845A. An example of the diode
2. Select the test voltage and current appropriate for the diode being tested by pressing
3. Connect the test leads as shown in Figure 4-8.
Function modifiers:
None
test display is shown below.
the associated soft keys.
F1
+
Figure 4-8. Diode Testing Connections
0.5994
F2
F3
F4
F5
2
2W/4W
4
mA
1000 V CAT I
00
HI
LO
600V CAT II
V
INPUT
FUSED
REAR
1V
SENSE
FRONT
4 W
300V
10A
HI
LO
caw024.eps
caw13f.eps