CUB4TC00 Red Lion Controls, CUB4TC00 Datasheet - Page 4

METER PANEL LCD THERMOCOUPLE

CUB4TC00

Manufacturer Part Number
CUB4TC00
Description
METER PANEL LCD THERMOCOUPLE
Manufacturer
Red Lion Controls
Series
CUB4r
Type
Thermocoupler
Datasheet

Specifications of CUB4TC00

Display Type
LCD Non-Backlit
Display Face Size
2.95" L x 1.54" W (74.9 x 39.1mm)
Display Digits
5
Display Digits - Height
0.480" (12.19mm)
Backlight
Without
Mounting Type
Panel Mount
Termination
Screw Terminals
Voltage - Supply
9 V ~ 26 V
Applications
Thermocouple Input - Type T, E, J, K, R, S, B, N, or mV
Number Of Digits/alpha
5
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
Q1421239
Q763911
CALIBRATING THE CUB4RT
accurate temperature measurements. Over time, the electrical characteristics of
the components inside the CUB4RT will slowly change with the result that the
stored calibration values no longer accurately define the input circuit. For most
applications, recalibration every 1 to 2 years should be sufficient.
calibration. Allow 30 minute warm-up before performing any calibration related
procedure. The following procedures should be performed at an ambient
temperature of 15 to 35°C (59 to 95°F).
calibrating electronic equipment.
CAUTION: The accuracy of the calibration equipment will directly affect the
dISP
number or tenths. Tenths is not available for thermocouple types R, S, and B.
choices. Press the “SEL” button to enter the choice and return to the main menu.
FLtEr
variations or noise, increased levels of filtering will help to stabilize the display.
Software filtering effectively combines a fraction of the current input reading
with a fraction of the previous displayed reading to generate the new display.
for the filter uses 1/4 of the new input and 3/4 of the previous display to generate
the new display. A filter value of 2 uses 1/8 new and 7/8 previous. A filter value
of 3 uses 1/16 new and 15/16 previous.
choices. Press the “SEL” button to enter the choice and return to the main menu.
rANGE
applied to the new input sample. If the new input value is within the ± range of
the previous display, the filter will be applied to the new input.
(885°C), so the filter is applied to the new input signal.
response to large signal changes and maximum filtering of small signal changes.
digit. To select a new digit, press the “SEL” button. After the desired value is
on the display, press and hold the “SEL” button for three seconds to enter the
value and return to the main menu.
The CUB4RT uses stored voltage and resistance calibration values to provide
Calibration of the CUB4RT involves a voltage calibration and a resistance
Calibration should only be performed by individuals experienced in
accuracy of the CUB4RT.
This program item sets the display for Fahrenheit or Celsius and whole
To change the display, use the “blank” button to scroll through the list of
This menu item does not appear if
If the displayed process signal is difficult to read due to small process
Filter values represent no filtering (0), up to heavy filtering (3). A value of 1
To select the filter level, use the “blank” button to scroll through the list of
The Range value selected determines if the filter, as set previously, will be
Example:
885 ± 10 = 875 to 895 degree filter range based on previous display.
The new input sample (891 °C) is within ± 10 °C of the previous display
If the new input is outside the filter range, no filter is used. This allows fast
To enter the range value, use the “blank” button to increment the flashing
new input sample = 891 °C
previous display = 885 °C
-
-
-
0C
0
00000
,
,
1
0. 0 C
,
range = 10 °C
2
,
to
,
3
0F
00099
,
0. 0 F
tYPE
=
LIN
.
4
Calibration Check
1. Connect a precision resistance decade box with an accuracy of
2. Verify that the unit is programmed for the correct RTD type.
3. Verify that the unit is programmed for a value of zero for
4. Verify that the input jumper is in the correct position.
5. Set the precision resistance decade box to the values shown below and verify
OFSEt
display. Typically this value should be zero. Changing this value does not affect
calibration, but will affect the displayed temperature. If a constant offset error is
occurring in the application, this feature may be used to reduce or eliminate
that error.
or the sign of the number. To select a new digit or sign of the number, press the
“SEL” button. If no digits are flashing, use the “blank” button to toggle the sign
of the number or press the “SEL” button to wrap around to the LSD. After the
desired value is on the display, press and hold the “SEL” button for three
seconds to enter the value and return to the main menu.
CAL
section before attempting to calibrate this unit. If calibration is entered by
accident, press and hold the “SEL” button when the display shows
access code. Approximately three seconds later the display will return to
in the Program Menu.
YES
Jumper
to the unit as a three wire RTD.See Calibration Wiring Diagram All RTD
Wires should be the same length, gauge, and material. Keep wire length to a
minimum.
that the display reads accordingly.
other RTD types.
The offset value is used to add or subtract a constant temperature from the
To enter an offset value, use the “blank” button to increment the flashing digit
This item places the unit in calibration mode. Read the entire Calibration
To enter the Calibration Mode, use the “blank” button to change the
Type
. Press the “SEL” button to enter the choice.
-
YES
- ±
OHMS
140
200
260
320
380
396
/
17
20
80
00000
NO
calibrate the unit. Access codes are contained in the calibration
procedures. This unit is calibrated from the factory. Initial
calibration is not necessary. Calibration should only be performed
by individuals experienced in calibrating electronic equipment.
Pos B
Pt392
Read the entire Calibration instructions before attempting to
-200.1
-193.2
102.1
261.4
429.3
607.1
797.3
850.3
-49.9
°C
to
99999
OHMS
140
200
260
320
380
391
18
20
80
Pos B
Pt385
-201.1
-196.5
104.0
266.4
437.9
619.9
815.2
852.5
-50.8
°C
0.9 C for Cu427, and
OHMS
120
170
220
270
320
370
379
66
70
Pos B
Ni672
120.7
170.1
213.8
252.5
259.0
-80.9
-74.8
64.9
-0.0
°C
OFSEt
OHMS
10
12
14
16
18
20
5
6
8
0.7 C for all
.
Cu427
Pos A
0.01 ohms
00000
-103.2
NO
128.5
180.1
231.3
282.6
-78.0
-26.8
25.0
76.7
°C
tYPE
to a
for