CUB4TC00 Red Lion Controls, CUB4TC00 Datasheet - Page 6

METER PANEL LCD THERMOCOUPLE

CUB4TC00

Manufacturer Part Number
CUB4TC00
Description
METER PANEL LCD THERMOCOUPLE
Manufacturer
Red Lion Controls
Series
CUB4r
Type
Thermocoupler
Datasheet

Specifications of CUB4TC00

Display Type
LCD Non-Backlit
Display Face Size
2.95" L x 1.54" W (74.9 x 39.1mm)
Display Digits
5
Display Digits - Height
0.480" (12.19mm)
Backlight
Without
Mounting Type
Panel Mount
Termination
Screw Terminals
Voltage - Supply
9 V ~ 26 V
Applications
Thermocouple Input - Type T, E, J, K, R, S, B, N, or mV
Number Of Digits/alpha
5
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
Q1421239
Q763911
VOLTAGE CALIBRATION
1. Connect a precision Voltage Source with an output range of -10.000 to 60.000
2. Enter the Program Menu and step through the Program Menu list by pressing
3. Press the “blank” button to change
4. The display shows
5. After entering the correct access code, the unit will display
6. After voltage calibration is complete, the unit will return to the
7. Repeat calibration if necessary.
mV and an accuracy of 0.03% or better to the TC- (#5) and TC+ (#3)
terminals. Cold junction calibration must be performed AFTER voltage
calibration.
the “SEL” button until
code to
table below for calibration sequence.The unit displays
seconds after pressing the “SEL” button before proceeding to the next input
point.
item. Exit programming and verify various points from -10.00 to 60.00
mVDC using the Voltage Calibration Check procedure to ensure that the
calibration is correct.
DISPLAY
IN1
IN2
IN3
IN4
IN5
IN6
IN7
IN8
00006.
Press and hold “SEL”.
PARAMETER
-10.000 mV
10.000 mV
20.000 mV
30.000 mV
40.000 mV
50.000 mV
60.000 mV
0.000 mV
00000.
Voltage Calibration Wiring
CAL
is displayed.
Use the “blank” button to increment the access
Apply -10.000 mV, wait 5 sec., press
Apply 10.000 mV, wait 5 sec., press
Apply 20.000 mV, wait 5 sec., press
Apply 30.000 mV, wait 5 sec., press
Apply 40.000 mV, wait 5 sec., press
Apply 50.000 mV, wait 5 sec., press
Apply 60.000 mV, wait 5 sec., press
no
Apply 0.000 mV, wait 5 sec., press
to
YES
. Press “SEL”.
ACTION
CALC
IN1
CAL
. Follow the
for about 3
program
SEL
SEL
SEL
SEL
SEL
SEL
SEL
SEL
6
COLD JUNCTION CALIBRATION
The ambient temperature at which this calibration should be performed is 15 to
35 °C (59 to 95 °F). A Temperature indicator with 0.1 °C resolution and
accuracy is required to perform this calibration.
1. Enter Programming Menu, and verify the following:
2. Connect a thermocouple probe of known accuracy and type (J, K, E, T, or N)
3. Connect a reference temperature probe to measuring end of the CUB4TC
4. Compare the unit display with reference temperature indicator (or calibration
5. Enter the Program Menu and step through the Program Menu list by pressing
6. Press the “blank” button to change
7. The display shows
8. After entering the correct access code, the display will alternate between
9. If the current CJC value is 0.0 °C or 59.9 °C, set the
10. Otherwise, the new value to be entered is the sum of the current value and
11. Exit the Programming Menu and repeat the calibration from Step 4 to verify
This procedure must be performed AFTER an accurate voltage calibration.
Make any necessary changes to programming.
to the CUB4TC. The probe should match the one selected in programming.
thermocouple.The two probes should be shielded from air movement and
allowed sufficient time (15 minutes minimum) to equalize in temperature.
(As an alternative, the CUB4TC thermocouple probe may be placed in a
calibration bath of known temperature.)
bath). If the displayed unit temperature does not equal the reference probe
temperature, calculate the CJ error as follows:
CJ Error = reference probe temperature - CUB4TC displayed temperature
the “SEL” button until
code to
and the current cold junction compensation value.
ambient temperature, exit calibration, and repeat the CJC calibration
procedure.
the CJ Error recorded previously.
accuracy.
00003.
Press and hold “SEL”.
00000.
Cold Junction Calibration Wiring
OFSEt
tyPE
dISP
CJC
CAL
= probe being connected to the unit.
=
=
=
is displayed.
Use the “blank” button to increment the access
YES
0. 0 °C
0
no
to
YES
. Press “SEL”.
CJC
value to the current
CJC