FLUKE-190-204/AM/S Fluke Electronics, FLUKE-190-204/AM/S Datasheet - Page 115

no-image

FLUKE-190-204/AM/S

Manufacturer Part Number
FLUKE-190-204/AM/S
Description
SCOPEMETER 4CH 200MHZ COLOR SCC
Manufacturer
Fluke Electronics
Series
ScopeMeter® 190r
Type
Handheldr

Specifications of FLUKE-190-204/AM/S

Bandwidth
200MHz
Channels
4
Display Type
LCD - Color
Features
DMM, USB Interface, Save, Recall, Record, Trend
Probe Type
10:1 Voltage Probe (1 Blue, 1 Green, 1 Red, 1 Gray)
Sampling Rate (per Second)
1.25GS/s
Input Impedance
1M
Rise Time (typ)
1.7nS
Voltage - Input (max)
CAT III 1000V, CAT IV 600V
Voltage - Supply
7.2V Battery, 17.8VDC Adapter
Includes
Alligator Clips, Battery Charger/Adapter, Case, Hook Clips, Probe Set, Software, Test Leads, USB Cable
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
614-1175
Troubleshooting
The Test Tool Shuts Down After a Short Time
The Screen Remains Black
‘Setting the Power Down timer’.
Press
The batteries may be empty. Check the battery
symbol at the top right of the screen. A
indicates that the batteries are empty and must be
charged. Connect the BC190 power adapter.
The test tool is still on but the ‘display auto off’ timer is
active, see Chapter 6 ‘Setting the Display AUTO-off
Timer’. To turn the display on press any key (restarts
the ‘display AUTO-off’ timer), or connect the BC190
power adapter.
The power down timer is active, see Chapter 6
Make sure that the test tool is on (press
You might have a problem with the screen contrast.
Press
the arrow keys to adjust the contrast.
USER
to turn the test tool on.
, then press
F4
. Now you can use
symbol
)
The Test Tool Cannot Be Turned Off
If the test tool cannot be turned off due to a software
hangup, do the following:
FlukeView Does Not Recognize The Test Tool
The display ‘auto off’ timer is active, see Chapter 6
‘Setting the Display AUTO-off Timer’. To turn the
display on press any key (restarts the ‘display AUTO -
off’ timer), or connect the BC190 power adapter.
Keep the ON/OFF key pressed for at least 5 seconds.
Make sure that the test tool is turned on.
Make sure that the interface cable is properly
connected between the test tool and the PC. Use only
the test tool mini USB port for communication with a
computer!
Make sure that no SAVE/RECAL/COPY/MOVE action
from or to a USB-stick is being performed.
Make sure that the USB driver has been correctly
installed, see Appendix A.
Troubleshooting
107