FLUKE-190-204/AM/S Fluke Electronics, FLUKE-190-204/AM/S Datasheet - Page 72

no-image

FLUKE-190-204/AM/S

Manufacturer Part Number
FLUKE-190-204/AM/S
Description
SCOPEMETER 4CH 200MHZ COLOR SCC
Manufacturer
Fluke Electronics
Series
ScopeMeter® 190r
Type
Handheldr

Specifications of FLUKE-190-204/AM/S

Bandwidth
200MHz
Channels
4
Display Type
LCD - Color
Features
DMM, USB Interface, Save, Recall, Record, Trend
Probe Type
10:1 Voltage Probe (1 Blue, 1 Green, 1 Red, 1 Gray)
Sampling Rate (per Second)
1.25GS/s
Input Impedance
1M
Rise Time (typ)
1.7nS
Voltage - Input (max)
CAT III 1000V, CAT IV 600V
Voltage - Supply
7.2V Battery, 17.8VDC Adapter
Includes
Alligator Clips, Battery Charger/Adapter, Case, Hook Clips, Probe Set, Software, Test Leads, USB Cable
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
614-1175
Fluke 190 series II
Users Manual
Triggering on External Waveforms
(models 190-xx2)
Use external triggering when you want to display
waveforms on inputs A and B while triggering on a third
signal. You can choose external triggering with automatic
triggering or with edge triggering.
In this example you continue from the Trigger on Edges
example. To choose the external signal as trigger source,
continue as follows:
64
1
2
3
Supply a signal to the red and black 4-mm
banana jack inputs.
F1
TRIGGER
Display the
key labels.
Select Ext (external) edge trigger.
TRIGGER
(On Edges)
Observe that the key labels at the bottom of the screen
have been adapted to allow selection of two different
external trigger levels: 0.12 V and 1.2 V:
From this point the trigger level is fixed and is compatible
with logic signals.
4
F3
Select 1.2V under the Ext LEVEL
label.