85134F AGILENT TECHNOLOGIES, 85134F Datasheet - Page 43

Cable Sets

85134F

Manufacturer Part Number
85134F
Description
Cable Sets
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 85134F

Kit Contents
2 Cables
Peak Reflow Compatible (260 C)
No
Cable Length
24.75"
Leaded Process Compatible
No
Connector Type B
3.5mm Female
Connector Type A
2.4mm Female
Lead Length
62.9cm
Lead Free Status / RoHS Status
Contains lead / RoHS non-compliant
Table A-1 Connector Care Quick Reference
85134E/F/H & 85135E/F
Do
Keep connectors clean
Extend sleeve or connector nut
Use plastic end-caps during storage
Do
Inspect all connectors carefully
Look for metal particles, scratches, and dents
Do
Try compressed air first
Use isopropyl alcohol
Clean connector threads
Do
Clean and zero the gage before use
Use the correct gage type
Use correct end of calibration block
Gage all connectors before first use
Do
Align connectors carefully
Make preliminary connection lightly
Turn only the connector nut
Use a torque wrench for final connect
a. Use isopropyl alcohol in a well-ventilated area, allowing adequate time for moist
alcohol to evaporate and fumes to disperse prior to energizing equipment.
a
Handling and Storage
Making Connections
Connector Cleaning
Gaging Connectors
Visual Inspection
Twist or screw any connection
Do Not
Touch mating-plane surfaces
Set connectors contact — end down
Do Not
Use a damaged connector — ever
Do Not
Use any abrasives
Get liquid into plastic support beads
Do Not
Use an out-of-spec connector
Do Not
Apply bending force to connection
Over tighten preliminary connection
Tighten past torque wrench “break” point
Principles of Microwave Connector Care
Connector Care Quick Reference
A-3