FLUKE-190-104/AM Fluke Electronics, FLUKE-190-104/AM Datasheet - Page 33

no-image

FLUKE-190-104/AM

Manufacturer Part Number
FLUKE-190-104/AM
Description
SCOPEMETER 4CH 100MHZ COLOR AMER
Manufacturer
Fluke Electronics
Series
ScopeMeter® 190r
Type
Handheldr

Specifications of FLUKE-190-104/AM

Bandwidth
100MHz
Channels
4
Display Type
LCD - Color
Features
DMM, USB Interface, Save, Recall, Record, Trend
Probe Type
10:1 Voltage Probe (1 Blue, 1 Green, 1 Red, 1 Gray)
Sampling Rate (per Second)
1.25GS/s
Input Impedance
1M
Rise Time (typ)
3.5nS
Voltage - Input (max)
CAT III 1000V, CAT IV 600V
Voltage - Supply
7.2V Battery, 17.8VDC Adapter
Includes
Alligator Clips, Battery Charger/Adapter, Case, Hook Clips, Probe Set, Software, Test Leads, USB Cable
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
614-1176
See also Table 2 in Chapter 8.
Selecting AC-Coupling
After a reset, the test tool is dc-coupled so that ac and dc
voltages appear on the screen.
Use ac-coupling when you wish to observe a small ac
signal that rides on a dc signal. To select ac-coupling, do
the following:
Observe that the bottom left of the screen displays the
ac-coupling icon:
You can define how Auto Set affects this setting, see
Chapter 6 ‘Changing the Auto Set Options’.
1
2
4
F2
F4
A
.
Exit the menu
Display the
Highlight
AC
INPUT A
.
key labels.
Reversing the Polarity of the Displayed
Waveform
To invert, for example the input A waveform, do the
following:
For example, a negative-going waveform is displayed as
positive-going waveform which may provide a more
meaningful view. An inverted display is identified by an
inversed trace identifier (
and in the status line below the waveform.
1
2
3
4
F4
F4
A
ENTER
Display the
Open the
Select Inverted and accept
inverted waveform display.
Exit the menu.
) at the right of the waveform,
Acquiring Waveforms
INPUT A
INPUT A
menu.
key labels.
25