FLUKE-190-104/AM Fluke Electronics, FLUKE-190-104/AM Datasheet - Page 37

no-image

FLUKE-190-104/AM

Manufacturer Part Number
FLUKE-190-104/AM
Description
SCOPEMETER 4CH 100MHZ COLOR AMER
Manufacturer
Fluke Electronics
Series
ScopeMeter® 190r
Type
Handheldr

Specifications of FLUKE-190-104/AM

Bandwidth
100MHz
Channels
4
Display Type
LCD - Color
Features
DMM, USB Interface, Save, Recall, Record, Trend
Probe Type
10:1 Voltage Probe (1 Blue, 1 Green, 1 Red, 1 Gray)
Sampling Rate (per Second)
1.25GS/s
Input Impedance
1M
Rise Time (typ)
3.5nS
Voltage - Input (max)
CAT III 1000V, CAT IV 600V
Voltage - Supply
7.2V Battery, 17.8VDC Adapter
Includes
Alligator Clips, Battery Charger/Adapter, Case, Hook Clips, Probe Set, Software, Test Leads, USB Cable
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
614-1176
You will see a screen that looks like Figure 11.
Observe that the top right of the screen displays
SPECTRUM.
If it displays LOW AMPL a spectrum measurement cannot
be done as the waveform amplitude is too low.
3
4
5
6
ENTER
ENTER
ENTER
ENTER
Jump to Waveform: and select
Mathematics... to open the
Mathematics menu.
Select Function: Spectrum.
Select the source waveform for the
spectrum: Source : A, B, C or D
Select Window: Auto (automatic
windowing), Hanning, Hamming,
or None (no windowing).
If it displays WRONG TB the time base setting does not
enable the test tool to display an FFT result. It is either too
slow, which can result in aliasing, or too fast, which results
in less than one signal period on the screen.
10
7
8
9
F1
F2
F3
F4
Perform a spectrum analysis on
trace A, B, C or D.
Set the horizontal amplitude scale
to linear or logarithmic.
Set the vertical amplitude scale to
linear or logarithmic.
Turn the spectrum function off/on
(toggle function).
Acquiring Waveforms
29