U3401A-ABA AGILENT TECHNOLOGIES, U3401A-ABA Datasheet - Page 107

no-image

U3401A-ABA

Manufacturer Part Number
U3401A-ABA
Description
English Users And Service Guide 3400
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U3401A-ABA

Rohs Compliant
NA
For Use With
U3400 Series U3401A Dual Display Multimeter
Range
Table 6-5 ACI resolution, full scale reading, and burden voltage [±(% of reading + count)]
U3401A User’s and Service Guide
500 μA
5 mA
50 mA
500 mA
5 A
10 A
[1]
[2]
[3]
[4]
[5]
Typical at full range reading and voltage across the input terminals.
In 10 A range, >10-20 Aac is readable for 20 seconds maximum with audio warning.
Input Current < 3 A RMS.
Input Current >35 μA RMS.
Accuracy specified at input >5% of range and >1 A for 10 A range except other specified.
Resolution
10 nA
100 nA
1 μA
10 μA
100 μA
1 mA
AC Current (true RMS, AC Coupling mode)
Maximum
reading
510.00
5.1000
51.000
510.00
5.1000
20.000
[2]
30 Hz to 50 Hz 50 Hz to 2 kHz 2 kHz to 5 kHz 5 kHz to 20 kHz
1.5% + 50
1.5% + 40
1.5% + 40
1.5% + 40
2% + 40
2% + 40
[3]
[3]
Accuracy (One year; 23°C ± 5°C)
0.5% + 20
0.5% + 20
0.5% + 20
0.5% + 20
0.5% + 30
(<1kHz)
1.5% + 50
1.5% + 40
1.5% + 40
1.5% + 40
Specifications and Characteristics
Not Specified
[5]
3% + 60
3% + 75
3% + 60
3% + 60
[4]
Burden voltage
and shunt resistor
<0.06 V/100 Ω
<0.6 V/100 Ω
<0.08 V/1 Ω
<0.8 V/1 Ω
<0.3 V/0.01 Ω
<0.6 V/0.01 Ω
[1]
89
6