U3401A-ABA AGILENT TECHNOLOGIES, U3401A-ABA Datasheet - Page 44

no-image

U3401A-ABA

Manufacturer Part Number
U3401A-ABA
Description
English Users And Service Guide 3400
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U3401A-ABA

Rohs Compliant
NA
For Use With
U3400 Series U3401A Dual Display Multimeter
2
Figure 2-6 Resistance /continuity terminal connection and display
26
CAU T ION
Operations and Features
Performing Resistance/Continuity Measurements
Disconnect circuit power and discharge all high-voltage capacitors before measuring
resistance, or testing circuit continuity, to avoid damaging the multimeter or the
multimeter under test.
Measuring resistance
1 Press
2 Connect the red and black test leads to the input terminal as shown in
3 Probe the test points (by shunting the resistor) and read the display.
In autoranging mode, the multimeter automatically selects the appropriate
range and the measurement is displayed.
• Six ranges:
• Measurement method:
• Input protection:
Figure 2-
Resistance
6.
Test current
.
+
500.00 , 5.0000 k , 50.000 k , 500.00 k , 5.000 M ,
50.000 M ,
Two-wire ohms, open circuit voltage limited to <5 V
500 V on all ranges
U3401A User’s and Service Guide