Recommended Test Equipment
U3401A User’s and Service Guide
The test equipments recommended for the performance verification
procedures are listed below. If the exact instrument is not available,
substitute calibration standards of equivalent accuracy.
A suggested alternate method would be to use the Agilent 3458A 8½ - digit
digital multimeter to measure less accurate yet stable sources. The output
value measured from the source can be entered into the instrument as the
target calibration value.
Table 4-1 Recommended test equipments
Application
Zero calibration
DC voltage
DC current
Resistance
AC voltage
AC current
Frequency
Diode
Recommended Equipment
Shorting plug—Dual banana plug with
copper wire short between the two terminals
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Recommended Accuracy
Requirements
< 1/5 instrument 1 year spec
< 1/5 instrument 1 year spec
< 1/5 instrument 1 year spec
< 1/5 instrument 1 year spec
< 1/5 instrument 1 year spec
< 1/5 instrument 1 year spec
< 1/5 instrument 1 year spec
Performance Test
59