U3401A-ABA AGILENT TECHNOLOGIES, U3401A-ABA Datasheet - Page 112

no-image

U3401A-ABA

Manufacturer Part Number
U3401A-ABA
Description
English Users And Service Guide 3400
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U3401A-ABA

Rohs Compliant
NA
For Use With
U3400 Series U3401A Dual Display Multimeter
6
94
Specifications and Characteristics
Measurement
Table 6-14 Supplemental measurement specifications
AC voltage
AC voltage (true
RMS, AC+ DC
coupling mode)
Measurement method
Crest factor
Input Impedance
Maximum input voltage
Overload ranging
Input protection
Response time
Measurement method
Crest factor
Input impedance
Maximum input voltage
Overload ranging
Input protection
Response time
Specification
AC coupled true RMS – measure the AC
component with up to 400 VDC bias on any range
Maximum 3:0 at full scale
1 MΩ in parallel with <100 pF of all ranges
1000 V RMS / 1400 V PEAK
2x10
input
1x10
input
Will select higher range if peak input overload is
detected during autoranging. Overload is reported
in manual ranging.
1000 V on all ranges
Approximate 1.5 second when the displayed
reading reaches 99.9% AC value of the tested
input signal at the same range.
AC coupled true RMS — measure the AC
component with up to 400 VDC bias on any range
Maximum 3:0 at full scale
1 MΩ in parallel with <100 pF
1000 V RMS / 1400 V PEAK
2x10
input
1x10
input
Will select higher range if peak input overload is
detected during autoranging. Overload is reported
in manual ranging.
1000 V RMS on all ranges
Approximate 1.5 second when the displayed
reading reaches 99.9% AC value of the tested
input signal at the same range.
7
6
7
6
V-Hz product on any range, common mode
V-Hz product on any range, normal mode
V-Hz product on any range, normal mode
V-Hz product on any range, common mode
U3401A User’s and Service Guide

Related parts for U3401A-ABA