CAU T ION
Performance Test
Table 4-2 DC voltage verification test
Set the calibrator output to 0 V before disconnecting it from the multimeter input
terminals.
DC Current Verification Test
1 Connect the calibrator to the front panel Hi and Lo input connectors.
2 Select each function and range in the order shown in
3 Make a measurement and observe the result. Compare measurement
Function
DC voltage
Provide the input shown in
results to the appropriate test limits shown in
allow for appropriate source settling when using the Fluke 5520A).
Input
0.00 V
0.0000 V
0.000 V
0.00 V
0.0 V
450.00 mV
4.5000 V
45.000 V
450.00 V
900.0 V
Table 4-
Range
500 mV
5 V
50 V
500 V
1000 V
500 mV
5 V
50 V
500 V
1000 V
3.
U3401A User’s and Service Guide
Table 4-
Error from nominal one year
±40 μV
±0.4 mV
±4 mV
±40 mV
±400 mV
±130 μV
±1.3 mV
±13 mV
±130 mV
±500 mV
Table 4-
3. (Be certain to
3.