U3401A User’s and Service Guide
Figure 2-1 ACV terminal connection and display
CAU T ION
Performing Voltage Measurements
Ensure that the terminal connections are connected correctly before making any
measurement. To avoid damaging the multimeter, do not exceed the rated input limit.
Measuring AC Voltage
1 Press
2 Connect the red and black test leads to the respective input terminals
3 Probe the test points and read the display.
In autoranging mode, the multimeter automatically selected the
appropriate range and the measurement is displayed.
• Five ranges:
• Measurement method:
• Crest factor:
• Input impedance:
• Input protection:
as shown in
ACV
voltage
source
+
.
Figure 2-
1.
500.00 mV, 5.00 V, 50.00 V, 500.00 V, 750.00 V
AC coupled true RMS—measure the AC component with up to
400 VDC bias on any range
Maximum 3:0 at full scale
1 MΩ±2%parallel with capacitance <100 pF on all ranges
750 V RMS on all ranges