U3401A-ABA AGILENT TECHNOLOGIES, U3401A-ABA Datasheet - Page 110

no-image

U3401A-ABA

Manufacturer Part Number
U3401A-ABA
Description
English Users And Service Guide 3400
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U3401A-ABA

Rohs Compliant
NA
For Use With
U3400 Series U3401A Dual Display Multimeter
6
Table 6-12 Range and accuracy (dB)
Voltage range
92
500 mV
5 V
50 V
500 V
1000 V (dc)
750 V (ac)
[1]
[2]
[3]
[4]
[5]
Autoranging is used when dBm operation is enabled.
In 1000 V (dc) range, 1200 V is readable. In 750 V (ac) range, 1000 V is readable.
Reading displayed in dB when Rel operation is used.
For input voltage at frequemcy between 50 Hz to 1 kHz.
Input Voltage <200 V RMS.
Specifications and Characteristics
[1][2]
Decibel Calculation
Sensitivity for Current Measurement
Table 6-11 Sensitivity for current measurement
Range
500 μA
5 mA
50 mA
500 mA
5 A
10 A
Input voltage
20 mV to 500 m V
500 mV to 5 V
5 V to 50 V
50 V to 500 V
500 V to 1000 V
500 V to 750 V
Minimum sensitivity (RMS Sine-wave)
30 Hz to 20 kHz
35 μA
0.25 mA
2.5 mA
25 mA
0.25 mA (<2 kHz)
2.5 A (<2 kHz)
dBm
600 Ω ref
–29.82 to –3.80
–3.80 to 16.20
16.20 to 36.20
36.20 to 56.20
56.20 to 62.22
56.20 to 59.72
[3]
range at
30 Hz to 50 Hz
Not Specified
0.2
Accuracy(One year; 23°C ± 5°C) (dB)
0.3
0.2
0.2
[5]
U3401A User’s and Service Guide
50 Hz to 10 kHz 10 kHz to 100 kHz
0.2
0.3
0.2
0.2
0.2
[4]
Not specified
0.5
0.7
0.5
0.5
[5]