10.1.5 HCS08-Specific: Communication Lost During Debugging
This problem may have several causes:
1.
2.
3.
10.1.6 HCS08-Specific: STOP Assembly Instruction Causes a
If the STOPE bit in the SOPT register is not set, the STOP instruction is
recognized as an illegal opcode, causing the microcontroller to reset. The
SOPT register can be written only once after reset.
10.2 Diagnostic Test
inDART-One has built-in self-test capabilities. This means that you can verify
by yourself, at any time, the correct operation of the instrument’s hardware.
To perform the diagnostic test:
1.
The microcontroller’s bus frequency has been changed by the user
application and the CLKSW bit is set to 1. In this case, set the CLKSW
bit to 0 (in the “Communication Settings” dialog box).
The BKGDPE bit in the SOPT (System Option Register) register has
been set to 0, configuring the BKGD pin as a generic I/O pin. The
BKGD pin must be reserved for the BDC communication (BKGDPE =
1).
A microcontroller reset has occurred. Among other things, this may be
caused by the COP peripheral. After reset, the COP is enabled, so your
program must either disable it or reset its timer.
Open the inDART-One Control Panel. Select “Start > Programs >
SofTec Microsystems > inDART-One > Control Panel”. The
following dialog box will appear.
Microcontroller Reset
inDART-One User's Manual
131