8808A 120V Fluke Electronics, 8808A 120V Datasheet - Page 47

DMM 5.5 RES RS-232 INTERFACE

8808A 120V

Manufacturer Part Number
8808A 120V
Description
DMM 5.5 RES RS-232 INTERFACE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8808A 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1068
Measuring Current
The Meter is capable of making ac and dc current measurements up to 10 A.
To perform a current measurement:
To avoid blowing the current fuse or damaging the Meter, do
not apply power to the circuit under test until test leads are
properly installed to the appropriate input terminals. For current
measurements above 200 mA, install the test leads into the 10 A
and LO terminals only.
1. Turn off power to the circuit under test.
2. Connect test leads between the Meter and the circuit under test.
3. If the circuit current is unknown, start by using the 10 A and LO terminals.
4. If the measurement is expected to be below 200 mA, connect the test leads to the
5. For measurements expected to be 200 mA to 10 A, connect the test leads to the
6. Press B to measure ac current or press E to measure dc current.
7. Apply power to the circuit under test.
Figure 3-7. Input Connections for 4-Wire Ohms Using 2x4 Wire Leads
200 mA and LO terminals only and remove any leads in the 10 A terminal. See
Figure 3-8.
10 A and LO terminals only. See Figure 3-9.
The Meter selects the appropriate range in autorange mode. The function and
measurement are displayed.
Ground tab
W Caution
Operating the Meter from the Front Panel
Selecting a Measurement Function
2W/4W
1000V
mA
750V
LO
HI
200 mA
MAX
MAX
V
INPUT
1000 V
600 V
500 V pk
10 A
MAX
1V
CAT I
CAT II
SENSE
4 W
300V
LO
10A
HI
eue26.eps