8808A 120V Fluke Electronics, 8808A 120V Datasheet - Page 54

DMM 5.5 RES RS-232 INTERFACE

8808A 120V

Manufacturer Part Number
8808A 120V
Description
DMM 5.5 RES RS-232 INTERFACE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8808A 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1068
8808A
Users Manual
3-20
Minimum / Maximum Modifier (MIN MAX)
To change the response level, press Q and I. The response level currently selected
(1, 2, 3 or 4) appears on the primary display. Press U or V to step to the desired
response level, then press R for two seconds to set the level and return to the primary
display. You can return to the primary display without changing the response level by
pressing any button except R, U or V.
The minimum/maximum (MIN MAX) modifier stores the minimum and maximum
inputs of your measurements.
Selecting the MIN MAX modifier turns off autoranging and locks in the present range, so
make sure you are in the correct range before pressing N. If you press U or V
after pressing N, the Meter exits the MIN MAX modifier mode.
To store the minimum and maximum inputs:
1. Press N to select the MIN MAX modifier.
2. Press N again to display the minimum reading. MIN is shown and the display
3. Press N again to display either the minimum or maximum reading. M
4. To exit MIN MAX mode, press and hold N for 2 seconds.
5. To observe the actual readings without resetting the stored values, press Q and
Level 1 (0.01 % of reading)
Level 2 (0.1 % of reading)
Level 3 (1 % of reading)
Level 4 (10 % of reading)
When N is first pressed, the minimum and maximum values are set to the
reading displayed. MAX is shown and the display indicates the latest maximum
reading.
indicates the latest minimum reading.
is shown and the display indicates either the minimum or maximum reading.
then select the same measurement function that you selected for the primary
display.