8808A 120V Fluke Electronics, 8808A 120V Datasheet - Page 70

DMM 5.5 RES RS-232 INTERFACE

8808A 120V

Manufacturer Part Number
8808A 120V
Description
DMM 5.5 RES RS-232 INTERFACE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8808A 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1068
8808A
Users Manual
4-10
Setting the Trigger Type Configuration
External Trigger via the Computer Interface
Slow
Medium
Fast
[1] Depends on A/D trigger speed.
[2] Depends on how fast the trigger signal is transmitted.
Rate
To set the trigger type configuration using the computer interface, enter the command
TRIGGER <type> (where <type> is the trigger type) and press Enter. See Table 4-3 for
trigger types.
Select trigger type 3 or 5 to enable the settling delay if the input signal is not stable before
a measurement is triggered. Typical settling delays are provided in Table 4-3. RS-232
reading transfer rates are provided in Table 4-4.
To trigger a measurement using the RS-232 computer interface, enter the command
*TRG and press Enter. See the “Common Commands” section later in this chapter for use
of *TRG command.
To trigger a measurement using pin 9 of the RS-232 interface, see
Figure 4-1.
In external trigger mode (mode 2 to mode 5), *TRG command is always
available.
If you enter the remote mode with trigger type 4 or 5 selected, the Meter remains in
its external trigger state; however, because the Meter is in the remote mode, you will
only be able to trigger measurements with rear trigger types 4 and 5. To exit remote
mode, perform steps 1 and 2 and select trigger type 2 or 3 (as applicable).
Internal Trigger Operation (Trigger 1)
Figure 4-1. External Trigger Using Pin 9 of RS-232 Interface
Table 4-4. RS-232 Reading Transfer Rates
+5 V Out (Pin 1)
Trig In (Pin 9)
100
2.5
20
[1]
[1]
[1]
Note
Readings per Second
External Trigger Operation (Trigger 4)
100
2.5
20
[2]
[2]
[2]
eue24.eps