8808A 120V Fluke Electronics, 8808A 120V Datasheet - Page 82

DMM 5.5 RES RS-232 INTERFACE

8808A 120V

Manufacturer Part Number
8808A 120V
Description
DMM 5.5 RES RS-232 INTERFACE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8808A 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1068
8808A
Users Manual
4-22
RANGE <value range>
RANGE1?
RANGE2?
RATE <speed>
RATE?
Command
Table 4-11. Range and Measurement Rate Commands and Queries (cont.)
Sets the primary display to <value range> where <value range> is the number
in the Range Value column of Table 4-11A that corresponds with the
applicable function ranges (voltage, ohms, current, etc.).
Returns the range presently selected on the primary display.
Returns the range presently selected on the secondary display. If the
secondary display is inactive, an Execution Error is generated.
Sets the measurement rate to <speed> where <speed> is either S for slow
(2.5 readings/second), M for medium (20 readings/second), or F for fast (100
readings/second).
S, M, and F can be entered as uppercase or lowercase. Any other entry for
<speed> generates an Execution Error.
Returns <speed> as S for slow (2.5 readings/second), M for medium (20
readings/second), or F for fast (100 readings/second).
[1] 1000 V dc, 750 V ac
Range
Value
1
2
3
4
5
6
7
200 mV
2 V
20 V
200 V
1000 V dc
NA
NA
Voltage
Range
Table 4-11A. Ranges for Each Function
[1]
200 e
2 ke
20 ke
200 ke
2 Me
20 Me
100 Me
Range
Ohms
Description
20 mA
200 mA
2 A
10 A
NA
NA
NA
Current
AC
2 kHz
20 kHz
200 kHz
1000 kHz
NA
NA
NA
Range
Freq.
200 µA
2000 µA
20 mA
200 mA
2 A
10 A
NA
Current
DC