8808A 120V Fluke Electronics, 8808A 120V Datasheet - Page 71

DMM 5.5 RES RS-232 INTERFACE

8808A 120V

Manufacturer Part Number
8808A 120V
Description
DMM 5.5 RES RS-232 INTERFACE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8808A 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1068
Status Registers
Event Byte Register
Service Request Enable
Register
Event Status Register
Event Status Enable
Register
Register
The contents of the status register (STB) are determined by the service enable register
(SRE), event status register (ESR), event status enable register (ESE), and the output
buffer. These status registers are explained in the following paragraphs and summarized
in Table 4-5.
Figure 4-2 shows the relationship of these registers.
Read Command
Table 4-5. Status Register Summary
*SRE?
*ESR?
*STB?
*ESE?
Operating the Meter Using the Computer Interface
Write Command
None
*SRE
None
*ESE
Enable Register
None
None
SRE
ESE
Status Registers
4-11
4