8808A 120V Fluke Electronics, 8808A 120V Datasheet - Page 85

DMM 5.5 RES RS-232 INTERFACE

8808A 120V

Manufacturer Part Number
8808A 120V
Description
DMM 5.5 RES RS-232 INTERFACE
Manufacturer
Fluke Electronics
Type
Digital (DMM)r
Datasheets

Specifications of 8808A 120V

Includes
Test Leads
Style
Bench
Display Digits
5.5
Display Type
VFD, Dual
Display Count
200000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, dB, Diode Test
Features
Hold, Min/Max, RS-232 Port
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Not applicable / RoHS non-compliant
Other names
614-1068
Miscellaneous Commands and Queries
RS-232 Remote / Local Configurations
^C (CONTRL C)
FORMAT <format>
FORMAT?
PRINT <rate>
SERIAL?
Volts dc
Volts ac
Amps dc
Amps ac
Resistance
Frequency
DIODE
Continuity Test
Command
Table 4-15 describes miscellaneous commands and queries.
Table 4-17 describes the RS-232 remote and local configuration commands, which are
used with the RS-232 interface to set up the remote/local configuration of the Meter.
These commands are valid only when the RS-232 interface is enabled.
Measurement Function
Table 4-16. Measurement Units Output with Format 2
Table 4-15. Miscellaneous Commands and Queries
Causes =><CR><LF> to be output.
Set output <format> to 1 or 2.
Format 1 outputs measurement values without measurement units (VDC,
ADC, OHMS, etc.).
Format 2 allows measurement units to be output with measurement units.
(See Table 4-16.) Format 2 primarily is used with RS-232 print-only mode.
Returns the format in use (1 or 2).
Sets print rate for print mode. See Table 4-2.
Returns Meter’s serial number.
Operating the Meter Using the Computer Interface
Description
Units Output
Computer Interface Command Set
OHMS
OHMS
VDC
ADC
VDC
VAC
AAC
HZ
4-25
4